Stainless Steel Care Instructions

Care instructions

304L stainless, treated right.

Bur blocks, hygiene cassettes, and the silicone inserts that sit inside them, taken through cleaning, drying, and sterilization without losing their finish.

All instructions, disclaimers, and warranties stated here apply to the entire order, its individual components, and all products available for purchase through dentps.com. Use or purchase of DentPS products constitutes agreement with the following recommendations and instructions.

Stainless grade

304L

Stainless steel bur blocks and hygiene cassettes.

Built for every process

Any

Ultrasonic baths, thermal disinfectors, chemiclaves, and autoclaves.

Ceiling

350°F / 176°C

Do not exceed this at any point in the process.

Refer to the instructions provided by the manufacturer of the cleaning or sterilizing machine you use. Their limits sit on top of ours.

Cleaning

  1. 01

    After use, wipe the instruments before storing them in the DentPS stainless steel holders.

  2. 02

    Clean the DentPS stainless steel holders, and the instruments in them, in enzymatic cleaner or potable water, using the thermal disinfector manufacturer's recommended solutions.

  3. 03

    Rinse thoroughly with distilled water afterward to remove detergent and particles.

  4. 04

    DentPS stainless steel products may be fitted with silicone inserts. Those inserts are also suitable for any decontamination, cleaning, and sterilization process.

  5. 05

    Dry completely. Complete drying before sterilization is not optional.

Sterilization

  1. 01

    Confirm all debris came off during cleaning.

  2. 02

    Pack the stainless steel holders in sterilization pouches or surgical wraps.

  3. 03

    Load the wrapped holders into the sterilizer and keep them slightly separated.

  4. 04

    Do not exceed 350°F / 176°C during any process.

  5. 05

    Follow the sterilizer manufacturer's instructions for operating the unit.

Recommendations

Water marks

Brownish marks on sterilized products are often mistaken for rust. They sit mainly where cleaning and rinsing cannot easily reach, hinges for example. Prevent them by rinsing thoroughly with reverse osmosis or distilled water and making sure everything is completely dry before sterilization.

To keep water marks off any stainless surface during sterilization, use distilled water only in the autoclave, since it carries no dissolved solids. Any other water leaves its solids behind as the instruments dry, and those solids are the marks you see on the instruments and the holder.

Corrosion

Disinfecting and ultrasonic cleaning baths left too long between solution changes concentrate as the water evaporates. If the instruments are not rinsed properly, those residues can burn brown during sterilization.

Infrequently, and only superficially, corrosion can come from contact with a strong acidic or caustic solution. It is usually rust carried over from instruments made of a lesser grade of stainless steel.

Questions

Send us the cycle and the mark.

If a holder comes out spotted, or you are unsure whether your sterilizer runs inside the ceiling, describe it and we will tell you what we would do.